Universal Tool Microscope(Image)19JPC-V
19LPC-V image-type all-purpose tool microscope is a new added image measuring system on the base of preserving all-purpose tool microscope performance. This system uses vision and special-purpose optical objective to substitute human’s eye to collimate and save video graph. Adopting electronic magnifier its zoom multiple is greatly improved and it is convenient to observe small work piece and precisely measure. Depending on high-precision track of all-purpose tool microscope, this system may realize high-precision image measuring in real meaning.
Image measuring parts optical objective lens
magnification |
Total magnification |
Objective view field |
Working distance(mm) |
3x |
25x |
Φ6 |
110 |
5x |
37x |
Φ4 |
95 |
Whole set of instrument
1. Mainframe: digital all-purpose tool microscope 19JC
2. Data gather
3. Planar measuring software
4. Brand computer and HP color ink-jet printer
5. Image measuring groupware (lifting arm, video-cam., and graph catcher)
6. Image measuring software
1. Main technical specification of universal tool microscope
Measuring range and division value X-Y-coordinate
Measuring range 200×100㎜ Division value 0.0005㎜
Collimation microscope Lifting distance:120mm
Arm of force inclining range: both 15°left and right Division value: 10′
Illumination grating adjusting range: φ3~φ32㎜ Division value:1㎜
Clinometers eyepiece Angle measuring range: 360° Division value: 1′
Figure eyepiece Angle measuring range: ±7° Division value:10′
Circulate arc recircle Curvature radius: R0.1~100㎜
Screw recircle Common screw distance t = 0.25 - 6㎜
Trapezium screw distance T = 2 - 20㎜
Optical indexing table Measuring range: 360° Division value:10″(projected value)
(Optional) Diameter of glass table φ106㎜
Optical dividing head Measuring range: 360°Division value:1′
Optical localizer Diameter of measuring headφ3±0.1㎜
(Actual high-point diameter error can not more than 0.5μm.)
Measuring force 8 -14g
Max. measuring depth 15㎜
Glass worktable Glass worktable dimension: 215×130㎜
Thimble bracket Max. clamping diameterφ100㎜
Max. clamping length: diameter of tested part≤ 55㎜,750mm
diameter of tested part > 55㎜,600mm
High thimble bracket Max. clamping diameter: φ180㎜
Max. Clamping length:600mm
V-type bracket:
Left V-type bracket front and behind adjusting range: Both5mmfront and behind.
Right V-type bracket upper and downward adjusting range: Upper15mm/ Downward3mm
Max. Carrying Capacity:40kg
2. Optical parameter of collimation microscope
Tag value of objective’s zoom multiple |
1x |
3x |
5x |
|
Total zoom multiple |
Clinometer or figure eyepiece |
10x |
30x |
50x |
Double-image eyepiece |
15x |
42x |
65x |
|
Field of view(mm) |
Clinometer or figure eyepiece |
φ20 |
φ6.6 |
φ4 |
Double-image eyepiece |
φ13
|
φ4.7 |
φ3 |
|
Working distance(mm) |
Clinometer or figure eyepiece |
81 |
90 |
65 |
Double-image eyepiece |
47 |
85 |
63 |
3. Instrument precision
Temperature requirement (1) Workroom temperature should be 20 ±2 ℃.
(2)Workroom temperature change can not exceed1 ℃per hour.
(3)Tested part and instrument temperature difference can not exceed0.5 ℃.
In the condition of meeting all temperature requirements, the instrument has the following guarantee:
(1)X-Y-coordination: When using glass millimeter ruler to examine, the Max. Un-precision of the instrument is (1 + L / 100)μm. L: Measuring length. Unit is mm
When the ruler corrects according to correction table, X-direction can not be more than 0.0035㎜. Y-direction can not be more than 0.0025.
(2) Clinometers eyepiece: Max. un-precision: not more than 1′.
(3) Double-image eyepiece: Instability of imaging: not more than 0.0005㎜
Incorrectness of imaging: not more than 0.001㎜.
(4)Optical dividing head: Max. un-precision: not more than 1′.
(5)Optical indexing table: Max. un-precision: not more than 30″ (optional).
(6)Optical graduator: Instability of measuring: not more than 0.001㎜.
Incorrectness of measuring: not more than 0.0015㎜
4. Weight of instrument and overall dimension
Mainframe net weight: 250kg
Overall dimension: (L ×W × W):980 ×1020 ×640 mm